Pulsed Xenon Irradiation Device
What is a Pulsed Xenon Irradiation Device?
A sterilization method using "light" as an alternative to sterilization methods using heat.
Because sterilization by "light" involves instantaneous emission, it is more effective than sterilization by heating.
- Contributes to reducing electricity costs
- Contributes to CO₂ reduction
- The object is less likely to deform or deteriorate.
- In-line sterilization is now possible, improving productivity.
Furthermore, by employing a xenon lamp with high germicidal capabilities,
- Because it uses mercury-free lamps, it is environmentally friendly and safe.
- Achieved a level of high-level sterilization (3-6D value).
An example of an object
- Infusion soft bag

- Injection syringe

- Vial

- Ampoule

- Plastic bottle caps

- Dessert container

- Bread

- Rice cake

Case studies
- Nissin Pharmaceutical Co., Ltd.

- Object: Physiological saline (20 ml plastic ampoule)
- Method of use: Final sterilization
- Otsuka Pharmaceutical Factory, Inc.

- Target materials: Physiological saline solution and drug solution (20 ml plastic ampoule)
- Method of use: Sterilization
- Toa Pharmaceutical Co., Ltd.

- Object: Chemical solution (150ml plastic bottle)
- Method of use: Final sterilization
- Industrial Research Institute of Niigata Prefecture Applied Materials Technical Support Center

- Target material: Nonwoven fabric for masks, etc.
- Usage: Performance testing
Examples of sterilization tests using pulsed light irradiation
Test conditions (2000J experimental machine)
- lamp
- Pulsed xenon lamps (2 lights) for 2000J
- Irradiation distance
- 75mm from the lamp center to the center of the bottle (from both sides)
- Charging energy
- 1000J, 1500J, 2000J
- Bacteria used
- Black koji mold: Aspergillus niger var. awamori (JCM2261)
- sample
- 20ml physiological saline (0.9%) PE bottle [Provided by Nisshin Pharmaceutical Co., Ltd.] *See photo in the bottom right
- Method for measuring bacterial count
- PE bottles injected with Aspergillus oryzae spores were pulsed at a fixed distance, then diluted to a fixed concentration, spread on PDA culture medium, and the bacterial count was measured.
Test Result
| Charging energy | Number of pulses required for 6D value of 20mℓ PE bottle |
|---|---|
| 1000J | 3 pulses |
| 1500J | 2 pulses |
| 2000J | 1 pulse |

Illuminance Monitoring System
In pharmaceutical-related factories, it is crucial to perform validation efficiently and accurately.